Catalog Number 999800756 |
Device Problems
Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Cyst(s) (1800); Edema (1820); Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191)
|
Event Date 11/26/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that the asr head and adaptor were removed and were replaced with 28mm head and competitor's polar dual mobility liner to remove the metal on metal articulation.Doi: unknown.Dor: (b)(6) 2019.Left hip.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. h6 patient code: no code available (3191) used to capture device revision or replacement.
|
|
Event Description
|
After review of medical records, the patient was revised for failure of left hip tha, metal on metal hip with elevated metal oins and metallosis.Post operative findings reported that the patient had a large effusion with abundant yellow fluid.Patient had ostelytic lession and had minimal wear on the trunnion.It was mentioned that the cup had no obvious wear.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.This hip replacement platform was voluntarily recalled from the market and the product codes are now considered inactive.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Search Alerts/Recalls
|