MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EXPRT REVISION KNEE; INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS Back to Search Results
Model Number 315-0C-717
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/10/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as poly exchange.The previous surgery and the surgery detailed in this event occurred 2 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to poly exchange.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - poly exchange.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRT REVISION KNEE
Common Device Name
INSERT, REVISION EXPRT, SZ C X 17MM, E-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key9938666
Report Number1644408-2020-00304
Device Sequence Number2078198
Product Code JWH
UDI-Device Identifier00888912114103
UDI-Public(01)00888912114103
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K140830
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Model Number315-0C-717
Device Catalogue Number315-0C-717
Device Lot Number013R1005
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2020
Initial Report FDA Received Date04/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-