Catalog Number 2N8378
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a clearlink system non-dehp catheter extension set that was attached to a patient's iv malfunctioned; further described as during computed tomography angiography the j loop split and resulted in a leak of contrast.Prior to the procedure, labs were taken and a j loop was attached and the set flushed well.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
Additional information was added: correction: the device was received inside a plastic biohazard bag for evaluation.The used sample had an unknown white solution in the fluid path and was contaminated with blood.A visual inspection was performed which observed a burst near the two piece luer lock assembly.Due to the nature of the returned sample, no additional tests were performed. the reported condition was verified.The cause of the condition could not be determined; however the probable cause was end user related.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key9938722
Report Number1416980-2020-02065
Device Sequence Number754691
Product Code FPA
UDI-Device Identifier00085412048918
UDI-Public(01)00085412048918
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K112893
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2N8378
Device Lot NumberUR20A29086
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2020
Initial Date Received by Manufacturer 03/13/2020
Supplement Date Received by Manufacturer04/28/2020
Initial Report FDA Received Date04/08/2020
Supplement Report FDA Received Date05/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age29 YR
-
-