| Model Number |
3L81-22 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Patient information: no further patient information was provided by the customer.An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated creatinine results generated on the architect c8000 processing module for a (b)(6) year old male patient.The following data was provided (reference range: 62 to 120 umol/l): on (b)(6) 2020; sid: (b)(6); initial result = 183.8 umol/l, repeat = 186.3 umol/l, repeat on cobas = 85.1 umol/l, repeat on (b)(6) 2020 = 75 umol/l.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of tickets was performed for lot number 20993un19.A review of tracking and trending did not identify any trends for the complaint issue.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the complaint issue.A review of labeling concluded that the issue is sufficiently addressed.Based on the investigation no systemic issue or deficiency of the creatinine assay for lot number 20993un19 was identified.
|
| |
|