| Model Number |
140.588 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Codes |
Pain (1994); Tissue Damage (2104)
|
| Date of Event |
02/07/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
After the evaluation was noticed that the item received is different from the item reported on guarantee form by the complainant.Therefore, the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.Moreover, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form and visual conference of the product returned by the dentist.
|
| |
|
Event or Problem Description
|
|
(b)(4)¿ the dentist reported that 2 months after the dental implant was installed in patient¿s mouth, its non-osseointegration was observed.Moreover, the patient presented bone type ii.
|
| |
|