Zimmer biomet complaint (b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2020-00202, 0001032347-2020-00203, 0001032347-2020-00204.Medical products: tmj system left narrow titanium mandibular component 45mm / 6 hole, part# 01-6546ti, lot# 907200.Tmj system left fossa component, small, part# 24-6563, lot# 933860c.Tmj system cross drive fossa screw 2.0mm x 7mm, part# 99-6577, lot# unk.2.4mm system 2.7mm x 10mm ht cross-drive screw,5/pk, part# 85-2710, lot# unk.The user facility is foreign; therefore, a facility medwatch report will not be available.Foreign report source: (b)(6).
|
|
This follow-up report is being submitted to relay additional information.The complaint is considered to be confirmed because a plan for a revision surgery was reported and a replacement custom tmj device was ordered.No product was returned; therefore, no functional tests or inspections could be performed.No x-rays, scans, pictures, or physician's reports were provided.The dhr for the mandible component was reviewed; no non-conformances were found.There are no indications of manufacturing defects.This is the only complaint for this item# 01-6546ti, lot# 907200.For all non-custom tmj mandibular implants in the previous one year (from the notification date) regarding revision due to trauma, there is a complaint rate of 0.07%, which is no greater than the occurrence listed in the application fmea.The most likely underlying cause of the complaint is the patient's reported accident.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|