The device was not returned to stryker sustainability solutions for evaluation.Customer stated the device is not allowed to be released at this time.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to the following: damage or separation to the white teflon pad or black pad replacement component(s).Too much force or torque applied to instrument, or grasping/pulling.Applying pressure between instrument blade and tissue pad without having tissue between them.Incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.The instructions for use (ifu) state: care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Clamping the tissue pad against the active blade without tissue on the full length of the blade will result in higher blade, clamp arm and distal shaft temperatures and can result in possible damage to the instrument.If this happens, there may be a system failure signaled by a continuous tone or alert screen when either of the foot pedals or hand control buttons is depressed.Keep the jaws of the device open when backcutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad and increased blade, clamp arm, and distal shaft temperatures.Thermal influences such as fluids or minimal to no tissue in the jaws may affect the presence or timing of the tone change.Blood and tissue buildup between the blade and shaft may result in abnormally high temperatures at the distal end of the shaft.To prevent burn injury, remove any visible tissue buildup at the distal end of the shaft.During and following activation in tissue, the instrument blade and clamp arm may become hot.Avoid unintended contact with tissue, drapes, surgical gowns, or other unintended sites at all times.Incidental and prolonged activation against solid surfaces, such as bone, may result in blade heating and subsequent blade failure, and should be avoided.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|
It was reported that during the case, the physician was using the device and the tip seemed to malfunction and separate.The device was reported to have visible charring and the white/clear portion of the tip appeared to have separated from the metal portion.Another device was pulled and functioned correctly.The case was able to continue and the patient was unharmed.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|