|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device powered up properly after the test battery was installed.Device also responded properly to button presses.No damage noted on the keypad traces or on the keypad dome switches.During visual inspection, the j2 keypad connector on the lcd/b was found locked properly.Device was monitored for 2 days.No charge/battery anomaly, unexpected low battery alarm, unexpected off no power alarm or unexpected e/a alarm noted.No drive/motor anomaly, unexpected motor noise anomaly or rewind anomaly noted.The motor was tested outside of the unit and passed.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|