| Model Number |
24716 |
| Medical Device Problem Codes |
Material Rupture (1546); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that balloon rupture and balloon detachment occurred.The target lesion was located in the heavily calcified left superficial femoral artery.A 6.0x220x150 sterling balloon catheter was advanced for dilatation.However, during the second inflation, the balloon ruptured circumferential and a piece of the balloon remained in the patient.The procedure took an extra three hours of trying to remove the piece of the balloon by snare and also a non-bsc filter wire that sheared off in the heavily calcified lesion.The fragmented piece had to be left in the patient and a covered stent was deployed to jail it up against the wall.The filter wire also tried to be retrieved and was unsuccessful, so it was jailed up against the wall along with the balloon fragment.There were no further complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture and balloon detachment occurred.The target lesion was located in the heavily calcified left superficial femoral artery.A 6.0x220x150 sterling balloon catheter was advanced for dilatation.However, during the second inflation, the balloon ruptured circumferential and a piece of the balloon remained in the patient.The procedure took an extra three hours of trying to remove the piece of the balloon by snare and also a non-bsc filter wire that sheared off in the heavily calcified lesion.The fragmented piece had to be left in the patient and a covered stent was deployed to jail it up against the wall.The filter wire also tried to be retrieved and was unsuccessful, so it was jailed up against the wall along with the balloon fragment.There were no further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast and blood in the inflation lumen and balloon.The balloon was loosely folded.There was a longitudinal tear 43mm long starting 100mm from the proximal end of the balloon.The balloon had a circumferential tear 143mm from the proximal end of the balloon.The inner shaft was separates at the tack weld.The distal end of the balloon and tip was missing.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Search Alerts/Recalls
|