Model Number FEM09060
Medical Device Problem Codes Positioning Failure (1158); Material Perforation (2205)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The lot number was provided; therefore, a lot history review is currently being performed.The sample was returned to bd for evaluation.The company is still investigating the issue at this time.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicates that model fem09060 endovascular stent graft experienced a positioning failure and material perforation.The information was received from a single source.The malfunction involved a patient with no reported patient injury.The (b)(6) year old female patient weighed (b)(6) lbs.
 
Additional Manufacturer Narrative
H10: h10: the initial mdr submitted for this malfunction inaccurately reported the mdr date of awareness.The correct mdr date of awareness is 03/31/2020.H10: the lot number was provided; therefore, a device history review was performed.The sample was returned to bd for evaluation and the alleged malfunction was confirmed.A definite root cause for the reported event could not be determined.The device is labeled for single use h10: g4 h11: g1,h6 (result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicates that model fem09060 endovascular stent graft experienced a positioning failure and material perforation.The information was received from a single source.The malfunction involved a patient with no reported patient injury.The 62 year old female patient weighed 232 lbs.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicates that model fem09060 endovascular stent graft experienced a positioning failure and material perforation.The information was received from a single source.The malfunction involved a patient with no reported patient injury.The 62 year old female patient weighed 232 lbs.
 
Additional Manufacturer Narrative
The initial mdr submitted for this malfunction inaccurately reported the mdr date of awareness.The correct mdr date of awareness is 03/31/2020.The lot number was provided; therefore, a lot history review is currently being performed.The sample was returned to bd for evaluation.The company is still investigating the issue at this time.The device is labeled for single use.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9943869
Report Number2020394-2020-02560
Device Sequence Number18999280
Product Code PFV
UDI-Device Identifier04049519008547
UDI-Public(01)04049519008547
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberFEM09060
Device Catalogue NumberFEM09060
Device Lot NumberANDR3721
Initial Date Received by Manufacturer 04/01/2020
Supplement Date Received by Manufacturer04/21/2020
06/19/2020
Initial Report FDA Received Date04/09/2020
Supplement Report FDA Received Date04/21/2020
06/19/2020
Patient Sequence Number1
-
-