• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GREENFIELD; FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GREENFIELD; FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Model Number 43175
Medical Device Problem Codes Break (1069); Positioning Problem (3009)
Health Effect - Clinical Codes Perforation (2001); Device Embedded In Tissue or Plaque (3165)
Date of Event 08/02/2019
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2011 the patient underwent placement of a greenfield vena cava filter.On 02 (b)(6) 2019, a computed tomography (ct) scan revealed a fractured filter, tilt and perforation.The proximal tip of the filter is located 9-10mm below the inferior margin of the right renal vein.The proximal apex of the filter is tilted anteriorly.Anterior margin of the apex of the filter attaching to the anterior wall of the inferior vena cava (ivc).Distal tips of all the peripheral struts of the filter appear to extend beyond the margins of the ivc in the axial plane.Right anterolateral strut at 10-11 oclock appears to course along the anterior ivc margin for at least 1cm.Anterior strut at 12 oclock also appears to course anterior to the ivc margin for at least 1.5cm.Left lateral strut at 3 oclock appears to course extraluminal for at least 2.5cm., distal tip noted close to the right posterolateral margin of the aorta, though not clearly invading the aorta.2 adjacent right posterolateral struts at 9 oclock and 8 oclock appear to be at least partially extraluminal distally for at least 5mm.This strut appears to be bent slightly posterolaterally towards the right, originally originating at the 6 oclock position.No obvious fracture of these struts.There are several clustered linear metallic foci/fragments in the left abdomen, anterior to the left kidney, which may represent surgical clips, which should be correlated clinically, as a piece of a fractured strut could potentially migrate into this region.
 
Event or Problem Description
On (b)(6) 2011 the patient underwent placement of a greenfield vena cava filter.On (b)(6) 2019, a computated tomography (ct) scan revealed a fractured filter, tilt and perforation.The proximal tip of the filter is located 9-10mm below the inferior margin of the right renal vein.The proximal apex of the filter is tilted anteriorly.Anterior margin of the apex of the filter attaching to the anterior wall of the inferior vena cava (ivc).Distal tips of all the peripheral struts of the filter appear to extend beyond the margins of the ivc in the axial plane.Right anterolateral strut at 10-11 oclock appears to course along the anterior ivc margin for at least 1cm.Anterior strut at 12 oclock also appears to course anterior to the ivc margin for at least 1.5cm.Left lateral strut at 3 oclock appears to course extraluminal for at least 2.5cm., distal tip noted close to the right posterolateral margin of the aorta, though not clearly invading the aorta.2 adjacent right posterolateral struts at 9 oclock and 8 oclock appear to be at least partially extraluminal distally for at least 5mm.This strut appears to be bent slightly posterolaterally towards the right, originally originating at the 6 oclock position.No obvious fracture of these struts.There are several clustered linear metallic foci/fragments in the left abdomen, anterior to the left kidney, which may represent surgical clips, which should be correlated clinically, as a piece of a fractured strut could potentially migrate into this region.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GREENFIELD
Common Device Name
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9944168
Report Number2134265-2020-04848
Device Sequence Number13596695
Product Code DTK
UDI-Device Identifier08714729114666
UDI-Public08714729114666
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K901659
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2014
Device Model Number43175
Device Catalogue Number43175
Device Lot Number0014253816
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/24/2020
Supplement Date Received by Manufacturer07/08/2020
Initial Report FDA Received Date04/09/2020
Supplement Report FDA Received Date07/22/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age57 YR
-
-