| Model Number |
109.664 |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Tissue Damage (2104)
|
| Date of Event |
10/04/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
In telephone contact, it was informed that the dentist did not have information about lot number of the product.Considering the surgical methodology, it was seen that the dentist made immediate load with a low torque (32n.Cm), witch is not indicated.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form and visual conference of the product returned by the dentist.
|
| |
|
Event or Problem Description
|
|
(b)(4) ¿ the dentist reported that 3 months after the dental implant was installed in intraoral region 22#, its non-osseointegration was observed.Moreover, 20n.Cm of primary stability was achieved and immediate load procedure was done.
|
| |
|