The device was received for evaluation.Evaluation determined that the device has a leak at the a rubber and a kink on the insertion tube.It was determined that a metal was protruding out of the a rubber.The device was placed for repair.As a precautionary measure, the instruction manual states the following; ¿do not twist or bend the bending section with your hands¿.Additionally, the instructions for safe use manual indicate that, damage to the bending section would happen if excessive force was applied while angulating in the opposite direction if there was no movement from the bending section; this would occur more so in a narrow environment.
|