| Model Number |
123526A |
| Medical Device Problem Code |
Partial Blockage (1065)
|
| Health Effect - Clinical Codes |
Thrombus (2101); No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient said he had blood clots that were clogging the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The reported event could not be confirmed.A potential failure mode could be ¿blocked lumen¿ with a potential root cause of ¿biological deposits¿.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿visually inspect the product for any imperfections or surface deterioration prior to use." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient said he had blood clots that were clogging the catheter.
|
| |
|