SMITH & NEPHEW, INC. LGN CONE IMPACTOR HANDLE; PRSTHSS,KN,PTLLFMRTBL,SM-CNSTRIND, CMNTD, PLYMR/MTL/PLYMR
|
Back to Search Results |
|
| Model Number |
71442011 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a revision surgery the impactor broke while impacting.It was outside the patient and any piece fell inside the patient.A smith and nephew backup device was available to complete the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.One of the locking tabs fractured off the device, rendering it inoperable.The fractured tab was not returned.The device was manufactured in 2017 and shows signs of extensive wear/usage.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.
|
| |
|