MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G34693
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant products: cook fs-qeb-a fusion quattro extraction balloon.Initial reporter; occupation: non-healthcare professional.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history records for the sphincterotome and wire guide lots were reviewed.The sphincterotome and wire guide device history records contain nonconformances that could potentially be related to wire guide damage.The device goes through different inspections prior to leaving the facility.These inspections would have removed any nonconforming devices prior to distribution.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Another possible contributing factor to wire guide damage is if the wire guide lumen is not flushed prior to wire guide advancement.Prior to distribution, all fusion® pre-loaded with acrobat wire guides are subjected to a visual inspection to ensure device integrity.A definitive cause for the reported observation could not be determined.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used an fs-omni fusion® pre-loaded with acrobat wire guide.After the sphincterotomy, the doctor used a fusion quattro extraction balloon.The balloon was stuck on the wire guide and it was impossible to move the balloon.The wire guide looped in the duodenum because of the 9 cm short wire guide outlet on the catheter.The doctor removed the wire guide and balloon and to start over the procedure with sphincterotome and wireguide and then balloon instead of fs-omni fusion® pre-loaded with acrobat wire guide.This extended the procedure time.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Common Device Name
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key9946238
Report Number1037905-2020-00171
Device Sequence Number3752123
Product Code KNS
UDI-Device Identifier10827002346939
UDI-Public(01)10827002346939(17)221217(10)W4298904
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K052051
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/17/2022
Device Model NumberG34693
Device Catalogue NumberFS-OMNI-ACRO-35-260
Device Lot NumberW4298904
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/18/2020
Initial Report FDA Received Date04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-