| Model Number |
2K8004 |
| Medical Device Problem Codes |
Defective Device (2588); No Pressure (2994)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Vyaire medical has received the defective component and results are pending completion of investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported while using the airlife adult manual resuscitator, an inability to ventilate.On (b)(6) 2020, the customer reported (b)(6) year old female from nursing home brought in due to choking airway cleared and was intubated (food visible below vocal cords).After exchanging the bags (manual resuscitator), the customer reported alternative ventilation was provided to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Results of investigation: the vyaire failure analysis laboratory received the suspect component for investigation.One sample was provided for the investigation.Sample was not inspected functionally because it was not decontaminated, only was inspected visually and found that the components snap ring and lim duck bill are disassembled from exhalation port.Therefore, defect reported was confirmed.This issue will be internally investigated within vyaire.
|
| |
|