MEDTRONIC NAVIGATION, INC STEALTHSTATION® S7; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9733857
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other relevant device(s) are: product id: 9733438, serial/lot #: (b)(4), ubd: 01-jan-2050, udi#: (b)(4).A medtronic representative went to the site to test the equipment.The issue was confirmed and the system camera, sensor control unit (scu), was replaced.The system then passed the system checkout and was found to be fully functional.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding an navigation system being used for a cranial resection procedure.During preparation, an "x" was displayed on the camera and all indicators did not illuminate.The camera details stated the "localizer was not connected".Since the lamp of the sensor control unit (scu) was not illuminated, the scu was replaced and the issue resolved.No further information was provided.
 
Additional Manufacturer Narrative
H3, h6: the scu 9733438 polaris camera (t301182) was returned for analysis.Analysis found that the returned scu powered up on the test bench with no issues.When connected to a known good psu the system had normal tracking with all tools.A check of the event log did not show any adverse events.No problem was found.Evaluation codes that apply to this testing: 10, 213, 67.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION® S7
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9948953
Report Number1723170-2020-01204
Device Sequence Number3435446
Product Code HAW
UDI-Device Identifier00613994450951
UDI-Public00613994450951
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 05/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9733857
Device Catalogue Number9733857
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2020
Supplement Date Received by Manufacturer05/26/2020
Initial Report FDA Received Date04/10/2020
Supplement Report FDA Received Date05/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-