| Model Number |
MMT-723NAB |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The device was not bumped or dropped prior to the alarm and the drive support cap was sticking out.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with detached end cap.Unable to perform functional testing including the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify a33 alarm due to detached end cap.Device received with loose/protruded drive support disk.
|
| |
|