| Model Number |
366999 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Three pantera leo balloon catheters were selected for pre-dilatation of a calcified lesion during the same intervention (20032003, 20032004 & 20032005).During inflation the three balloons ruptured at 14 to 16 bar.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned instrument was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed a fine jet of water emerging from the distal part of the balloon when attempting inflation.Microscopic analysis of the balloon surface showed a balloon pinhole just proximal to the distal x-ray marker and scratches on the balloon surface nearby the damage site.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy.
|
| |
|