Model Number 9600TFX26J
Device Problems Complete Blockage (1094); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Complete Heart Block (2627)
Event Date 02/27/2019
Event Type  Injury  
Manufacturer Narrative
Udi number: (b)(4).Per the instructions for use (ifu), conduction system defects (heart block) which may require a permanent pacemaker are potential adverse events associated with balloon aortic valvuloplasty, deployment of the prosthetic valve, and the overall tavr procedure.According to the valve academic research consortium (varc) guidelines, the close anatomical relationship between the aortic valve complex and the branching atrioventricular bundle may provide an explanation for these complications of the tavr procedure.According to literature review, and as documented in a technical summary written by edwards lifesciences, atrioventricular conduction disturbances after tavr are associated with many patient related and procedural related factors, including pre-operative co-morbid status, the degree and bulkiness of aortic valve and annular calcification, inter-ventricular septal thickness, pre-existing electrocardiogram abnormalities, the depth of prosthesis implantation, and the profile of the implanted prosthesis.Unlike conventional avr, where there may be localized trauma due to decalcification of the annulus and/or suture placement in the proximity of the av node or the bundles, tavr may cause conduction abnormalities through mechanical impingement of the conduction system by the prosthesis.The mechanisms of the development of heart block after tavr are well documented and described in the literature.It is also documented that pre-existing heart block is common in patients undergoing tavr or surgical avr and another 4-6 % will develop postoperative heart block, potentially requiring a permanent pacemaker.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.Patient factors not reported and/or the mechanisms described above are likely contributing factors for the event.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
As reported through the (b)(6) registry by our affiliates in (b)(6), post implant of a 26mm sapien 3 valve the patient developed complete av block that persisted.Permanent pacemaker implantation was performed on postoperative day (pod) 1 and the patient recovered.The valve was deployed in the aortic annulus 60:40 aortic/ventricular on the non-coronary cusp side and 80:20 aortic/ventricular on the right-coronary cusp side.Per the reporter, the valve deployment location was acceptable.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV)
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key9949704
MDR Text Key190043545
Report Number2015691-2020-11422
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/27/2020
Device Model Number9600TFX26J
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/16/2020
Initial Date FDA Received04/10/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received11/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-