• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET MICROFIXATION 1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK; PLATE, BONE Back to Search Results
Model Number N/A
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The user facility is foreign; therefore, a facility medwatch report will not be available.Report source - (b)(6).
 
Event or Problem Description
It was reported that the screws would not enter the bone during a cranial procedure.No adverse events were reported as a result of the malfunction.
 
Event or Problem Description
This follow-up report is being submitted to relay additional information.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The complaint is confirmed.A visual inspection found that the screw heads show signs of use and the tips of the screws show signs of wear from a failed insertion attempt.Functional testing was completed by attempting to insert the screws into a piece of white oak wood.The screws would not enter the wood when the insertion attempt was made.The dhr will not be reviewed as the lot number remains unknown.There are no indications of manufacturing defects.For this part (95-6104) and the previous one year (from the notification date) regarding the failure of insertion of the screws, (b)(4).The most likely underlying cause of the damage to the screws is the application of excessive force beyond what the screw is designed to encounter.The patient's bone density may have also been a factor.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.5MM SYSTEM HIGH TORQUE (HT),SD,X-DR,SCR,5/PK
Common Device Name
PLATE, BONE
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key9949928
Report Number0001032347-2020-00217
Device Sequence Number14267273
Product Code JEY
UDI-Device Identifier00841036056650
UDI-Public00841036056650
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121589
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number95-6104
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/03/2020
Supplement Date Received by Manufacturer08/24/2020
Initial Report FDA Received Date04/10/2020
Supplement Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-