| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.The device was put through extensive testing including bench handling and full functionality stress testing with known good electrode pads without duplicating the report.The analog board was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device displayed a "check pads" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|