|
The customer's report was observed during initial testing; however after disassembling the device to determine root cause, the malfunction was no longer seen.The device was put through extensive testing including bench handling, performing multiple 30 j tests, and ecg monitoring functional stress testing without duplicating the report.The analog board was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|