| Model Number |
MMT-523RNAH |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump alarmed compromised force sensor system.Customer stated drive support cap was sticking out.Customer stated they were unable to exit the preparing to prime loop.Troubleshooting was performed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but a33 alarm during the basic occlusion test due to protruded drive support disk.Unable to perform the occlusion, prime or a33 and excessive no delivery test due to a33 alarm.
|
| |
|