(b)(4).(b)(6).Visual inspection of the returned product found the blister seal to be intact.The foam inserts have been damaged such that foam debris are present within the blister.Visual inspection confirmed that the inner pouch containing the stem is torn such that the inner sterile barrier is compromised.Complaint sample was evaluated and the reported event was confirmed.Device history record was reviewed and no discrepancies were found.The likely condition of the part when it left zimmer biomet control is considered conforming based on the evaluation of the returned product, the dhr review, and the potential transit age of the device.The root cause for the presence of foam debris within the sterile cavity as well as the damaged inner pouch is attributed to damage during transit.This device falls within the scope of a corrective action which is to assess all current sterile barrier systems used to package products at zimmer biomet bridgend.As part of this corrective action, the pouch is being improved to use a stronger material (nylon), and foam end caps are being added.Also, the orientation the devices are packed in the shipper box is moving from vertical to horizontal, and the thickness of the shipper box has been increased.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|