MEDTRONIC SOFAMOR DANEK USA, INC ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 9960135
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the patient underwent posterior lumbar fusion revision.Intra-op, the instrument broke.No fragment of the broken product remained inside the patient.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
H6: product analysis: visually confirmed approximately 3mm of the instrument tip had been broken off and not was returned for analysis.Fracture surface analysis revealed a fairly flat ductile fracture with circular material flow, consistent with torsional overload, dimensional inspection of the tip diameter and material hardness confirmed conformance to print specification.This is consistent with torsional overload.Additional information: d10, g1, g2, h3, h6 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key9953709
Report Number1030489-2020-00445
Device Sequence Number2817457
Product Code LXH
UDI-Device Identifier00885074378329
UDI-Public00885074378329
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number9960135
Device Lot NumberOT11L006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2020
Initial Date Received by Manufacturer 03/16/2020
Supplement Date Received by Manufacturer06/26/2020
Initial Report FDA Received Date04/13/2020
Supplement Report FDA Received Date06/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-