| Model Number |
7010 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.The target lesion was located in the left anterior descending artery.A 3.5mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during the procedure, the catheter shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: returned product consisted of a quantum maverick balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There were numerous kinks.There was a complete separation at 68.2cm distal of the strain relief.There was contrast and blood in the inflation lumen.The balloon was tightly folded.Microscopic inspection revealed tip damage.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.The target lesion was located in the left anterior descending artery.A 3.5mm x 12mm quantum maverick balloon catheter was advanced for dilatation.However, during the procedure, the catheter shaft was fractured.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|