| Model Number |
N/A |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
Device was returned with packaging damage and the sterility barrier potentially compromised.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g4; g7; h1; h2; h3; h6.Visual inspection of the returned product identified that the box was opened and damaged and the sterile packaging exhibited stress marks and is cracked.The sterility of the product is compromised.Device history record (dhr) review identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the products when they left zimmer biomet control were conforming to specifications.The root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|