ZIMMER ORTHOPAEDIC MFG. LTD. FEMORAL COMPONENT OPTION SIZE C RIGHT COMPATIBLE WITH LPS-FLEX PROLONG; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
Device was returned with packaging damage and the sterility barrier potentially compromised.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g4; g7; h1; h2; h3; h6.Visual inspection of the returned product identified that the box was opened and damaged and the sterile packaging exhibited stress marks and is cracked.The sterility of the product is compromised.Device history record (dhr) review identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the products when they left zimmer biomet control were conforming to specifications.The root cause is attributed to transit damage.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL COMPONENT OPTION SIZE C RIGHT COMPATIBLE WITH LPS-FLEX PROLONG
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key9954725
Report Number3007963827-2020-00114
Device Sequence Number2764617
Product Code JWH
UDI-Device Identifier00889024001084
UDI-Public(01)00889024001084
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K173057
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number00596401352
Device Lot Number63075217
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/24/2020
Supplement Date Received by Manufacturer08/06/2020
Initial Report FDA Received Date04/13/2020
Supplement Report FDA Received Date08/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-