Model Number 8888123409P
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, it was noted that the catheter shaft (tube) had a leak just above the wings/hub.It was stated that the catheter was not repaired but was explanted and replaced.There was no reported patient injury.
 
Additional Manufacturer Narrative
Additional information: d10, g4, h3, h6 h3.Evaluation summary: post market vigilance (pmv) led a photographic evaluation of one device.The visual inspection of the returned photo noted: the image depicts the catheter with a red ring drawn around the area of the hub meeting the cannula.There is fluid in the cannula.The red and luer adapter clamps are locked at the extension tube site.There is no visible fluid in the extension tubes.Without the physical device functional evaluations are precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDROME
Common Device Name
CATHETER, HEMODIALYSIS, IMPLANTED
Manufacturer (Section D)
COVIDIEN MFG SOLUTIONS S.A.
edificio b20, calle #2
alajuela 20101
MDR Report Key9956523
Report Number3009211636-2020-00081
Device Sequence Number2764754
Product Code MSD
UDI-Device Identifier10884521157804
UDI-Public10884521157804
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123196
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8888123409P
Device Catalogue Number8888123409P
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2020
Initial Date Received by Manufacturer 03/31/2020
Supplement Date Received by Manufacturer04/29/2020
Initial Report FDA Received Date04/14/2020
Supplement Report FDA Received Date05/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-