| Model Number |
MMT-523NAS |
| Medical Device Problem Codes |
Failure to Cycle (1142); Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Power Problem (3010); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump received unexplained no delivery alarm, motor error,insulin pump error, diminished battery life, reservoir ring was crack as well as button was sticky.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Search Alerts/Recalls
|