• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR POROUS KNEE; DJO EMPOWR KNEETM, PRESS FIT BP, 5L

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR POROUS KNEE; DJO EMPOWR KNEETM, PRESS FIT BP, 5L Back to Search Results
Model Number 353-01-105
Health Effect - Clinical Codes Failure of Implant (1924); No Code Available (3191)
Date of Event 03/17/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as loosening / subsidence.The previous surgery and the surgery detailed in this event occurred 8.9 months apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were returned to manufacturer and the device evaluation anticipated but not yet begun at djo surgical.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision due to loosening/subsidence.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - revised left total knee for pain and possible loose/subsided tibial component.It was determined that the tibial component was not loose but it appeared to have possibly subsided into the tibia on the anterior and medial edges.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR POROUS KNEE
Common Device Name
DJO EMPOWR KNEETM, PRESS FIT BP, 5L
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx
MDR Report Key9957991
Report Number1644408-2020-00330
Device Sequence Number2438247
Product Code MBH
UDI-Device Identifier00190446230038
UDI-Public(01)00190446230038
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K171991
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/06/2024
Device Model Number353-01-105
Device Catalogue Number353-01-105
Device Lot Number190W1007A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/17/2020
Supplement Date Received by Manufacturer04/14/2020
Initial Report FDA Received Date04/14/2020
Supplement Report FDA Received Date05/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
341-10-705 LOT 065T1015
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
-
-