| Model Number |
27026BO |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to the factory for evaluation.As per their evaluation: upon evaluation it could be confirmed that the tip was loosened.The tip was returned; therefore, it is suspected that the tip was not left in patient.Remains of the adhesive are visible on the shaft and in the thread.The device was manufactured in january 2010.Lt is suspected that the device was in use for almost 10 years.The root cause most likely is end of device life due to wear and tear.No indication for a material or manufacturing related issue.
|
| |
|
Event or Problem Description
|
|
As per a report received by the factory in (b)(6), the tip of the obturator broke inside the patient and fell into the bladder.Tip was retrieved, and there was no harm to the patient.
|
| |
|