| Model Number |
2429-0007 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The devices have been received and the evaluation is pending.A follow up report will be submitted with investigation results once the evaluation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported from the outpatient surgery that the tip of the tubing broke off while spiking the bag of iv fluid.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer's report that the tip of the tubing broke off was confirmed.The drip chamber's spike tip was broken off.The broken off spike tip piece was not received for evaluation.Further inspection of the damaged spike surface observed the breakage to be rough and uneven.The nature of the breakage indicates a ductile fracture (vs.A brittle fracture) that is typically caused by an external impulse/impact force.No testing was deemed necessary due to the obvious breakage.The breakage was caused by an external lateral force.The root cause of the external lateral force was not identified.The device history record search for set model 2429-0007 lot 20015593 shows the set was manufactured on 07jan2020 with a total of (b)(4) units.There were no quality notifications issued during the production build of this set.
|
| |
|
Event or Problem Description
|
|
It was reported from the outpatient surgery that the tip of the tubing broke off while spiking the bag of iv fluid.There was no patient involvement.
|
| |
|