• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I OSSEOTITE 2 IMPLANT 3.75 X 11.5MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I OSSEOTITE 2 IMPLANT 3.75 X 11.5MM; DENTAL IMPLANT Back to Search Results
Model Number XFOS311
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/27/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).
 
Event or Problem Description
It was reported implant removed due to fracture.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).The osseotite® 2 implant 3.75 x 11.5mm (xfos311) was not returned.Since product has not been returned, visual/functional inspection could not be performed.The investigation has been performed based on the available information.No pre-existing conditions were noted on the per.The reported product was located on tooth # 34 (fdi) and used for approximately 4 years.The reported event could not be recreated due to the nature of the dental device and event (fracture).The customer has not provided additional pictures or x-ray images of the product.Dhr review was completed for the subject lot number 2015041092.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review: complaint history review was performed for the reported lot number (2015041092) for similar event and no other complaint was identified.Post market trend review: april post market trending was reviewed and there were no actionable events or corrective actions for the reported event (implant fracture) or product (xfos311).Therefore, based on the available information, device malfunction could not be verified and the reported event was non-verifiable.The following sections have been updated: b4: date of this report.B5: describe event or problem.G4: date received by manufacturer.G7: type of report, follow-up number.H2: follow up type.H3: device evaluated by manufacturer: change ¿no' to 'yes'.H6: evaluation codes.H10: additional narrative.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).One osseotite® 2 implant 3.75 x 11.5mm (xfos311) was returned for inspection.Review of the returned device notes that the implant is badly damaged and fractured across the threads.No pre-existing conditions were noted on the per.The reported product was located on tooth 34 (fdi) and used for approximately 4 years.The reported event could not be recreated due to the nature of the dental device and event (fracture).Dhr review was completed for the subject lot number 2015041092.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (2015041092) for similar event and no other complaint was identified.May post market trending was reviewed and there were no actionable events or corrective actions for the reported event (implant fracture) or product (xfos311).Therefore, based on the available information, device malfunction has occurred and the reported event was confirmed.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE 2 IMPLANT 3.75 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
MDR Report Key9961685
Report Number0001038806-2020-00684
Device Sequence Number14135235
Product Code DZE
UDI-Device Identifier00844868023791
UDI-Public(01)00844868023791(17)200429(10)2015041092(241)XFOS311
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K111216
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/29/2020
Device Model NumberXFOS311
Device Catalogue NumberXFOS311
Device Lot Number2015041092
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/27/2020
Supplement Date Received by Manufacturer06/05/2020
09/08/2020
Initial Report FDA Received Date04/15/2020
Supplement Report FDA Received Date06/11/2020
09/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-