• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VACCESS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VACCESS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number VA80104
Medical Device Problem Codes Deflation Problem (1149); Retraction Problem (1536)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
For the reported event, the device return to bd is pending.The photo review is currently underway review.The company is still investigating the issue at this time.The device was labeled for single use.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model va80104 pta balloon dilatation catheter allegedly experienced deflation problem and retraction problem.This report was received from a single source.This event did involve patient with no reported patient injury.The patients age, weight and gender were not provided.
 
Additional Manufacturer Narrative
H10: the lot number for the device was provided; therefore, a lot history review was performed.The device was not returned to the manufacturer for evaluation; however, a photo was provided.The investigation is inconclusive for deflation and retraction issue.The definitive root cause could not be determined.The device is labeled for single use.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
This report summarizes one malfunction.A review of the reported information indicated that model va80104 pta balloon dilatation catheter allegedly experienced deflation problem and retraction problem.This report was received from a single source.This event did involve patient with no reported patient injury.The patients age, weight and gender were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9962529
Report Number2020394-2020-02681
Device Sequence Number13573281
Product Code LIT
UDI-Device Identifier00801741121258
UDI-Public(01)00801741121258
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberVA80104
Device Catalogue NumberVA80104
Device Lot Number93NE0032
Initial Date Received by Manufacturer 03/31/2020
Supplement Date Received by Manufacturer06/26/2020
Initial Report FDA Received Date04/15/2020
Supplement Report FDA Received Date07/06/2020
Patient Sequence Number1
-
-