• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/30/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device was broken.A 3.0x40x90 (4f) sterling balloon catheter was selected for use.During preparation outside the patient, the device appeared to be frayed or disintegrated at the end of the wire.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr: returned product consisted of a sterling balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and balloon.The balloon was loosely folded.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that the device was broken.A 3.0x40x90 (4f) sterling balloon catheter was selected for use.During preparation outside the patient, the device appeared to be frayed or disintegrated at the end of the wire.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9962914
Report Number2134265-2020-05085
Device Sequence Number11515029
Product Code LIT
UDI-Device Identifier08714729860143
UDI-Public08714729860143
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K141112
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/03/2021
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0024884566
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Initial Date Received by Manufacturer 03/30/2020
Supplement Date Received by Manufacturer08/24/2020
Initial Report FDA Received Date04/15/2020
Supplement Report FDA Received Date08/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-