• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. UNI GLENOID-PERIPHERAL LOCK SCRW; PHX SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. UNI GLENOID-PERIPHERAL LOCK SCRW; PHX SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Model Number UNI GLENOID-PERIPHERAL LOCK SCRW
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
 
Event or Problem Description
It was reported that during implant to secured the baseplate, the ar-9145-36 screw broke off at the head.The distal end remains in the patient and the baseplate is secured with two other screws.The case was a reverse total shoulder.Update info 04/07/2020: the site was predrilled with an ar-9145k peripheral screw drill, 2.5 drill bit and a stryer drill.There was no torque indicator used.
 
Additional Manufacturer Narrative
Complaint confirmed, the screw broke in two, the returned fragment is 11mm long.Likely causes include improper bone prep, misaligned insertion, prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNI GLENOID-PERIPHERAL LOCK SCRW
Common Device Name
PHX SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
MDR Report Key9963761
Report Number1220246-2020-01794
Device Sequence Number14138291
Product Code PHX
UDI-Device Identifier00888867057739
UDI-Public00888867057739
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K142863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model NumberUNI GLENOID-PERIPHERAL LOCK SCRW
Device Catalogue NumberAR-9145-36
Device Lot Number18.01605
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/25/2020
Supplement Date Received by Manufacturer03/25/2020
Initial Report FDA Received Date04/15/2020
Supplement Report FDA Received Date04/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-