• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C4474
Device Problem Fire (1245)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/18/2020
Event Type  malfunction  
Manufacturer Narrative
Initial reporter address: (b)(6).The device was received for evaluation.Visual inspection revealed that the power cord was damaged.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the customer reported condition was determined to be the defective power cord.The power cord was replaced to resolve the issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the cable attached to the back of a homechoice device was on fire.This occurred during the last cycle of peritoneal dialysis therapy.The reporter further stated that the cable appeared to be melted.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park d
singapore 73875 0
SN   738750
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9964615
MDR Text Key187790831
Report Number1416980-2020-02190
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 04/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2020
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C4474
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/19/2020
Date Manufacturer Received03/18/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-