| Model Number |
OM-8010 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/30/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device, intended for use in treatment, was returned for evaluation.There was a relationship found between the device and the reported incident.Visual inspection shows a broken/detached grasper at distal end.During the functional evaluation both needles could be extracted by using a smart stich device because the pp-connector is a single use device.The pp-connector is broken.The complaint was verified and the root cause was determined to be a mechanical component failure.Corrective action has been taken to cover this issue.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|
Event or Problem Description
|
|
It was reported that the upper part of the mouth is broken off.Breakage occurred external to patient.Procedure was completed with a s and n back up device no delay or patient injury reported.
|
| |
|