| Model Number |
E053 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.The device emitted beeping tones associated to the code 1003.This device remains in service, but device replacement was recommended in less than 2 months since date of event.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned implantable cardioverter defibrillator (icd) was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device battery.Low voltage capacitors are used in the high voltage charging operation in order to facilitate fast charge times.The behavior of these capacitors resulted in a high current drain, which was depleting the device battery faster than normal.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis showed the battery appeared to be depleting more quickly than expected.The device emitted beeping tones associated to the code 1003.Device replacement was recommended.This device was already replaced and returned for analysis.Follow ups were increased as a result.No additional adverse patient effects were reported.
|
| |
|