• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPR 6MM HUM FRACT STEM PPS; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPR 6MM HUM FRACT STEM PPS; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Medical Device Problem Codes Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 03/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: catalog #: 113760, bio-mod hum head 40diax15mm, lot # 401330.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.The stem remains implanted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-01546.Will be returned.
 
Event or Problem Description
It was reported that the initial operation was performed approximately a month ago.Subsequently, the head implant disassociated from the stem making the joint dislocated a week later with no pain.A revision was performed two (2) days afterward.
 
Additional Manufacturer Narrative
(b)(4).G3: japan.Reported event was considered confirmed from the x-rays which stated that the hardware was disassociated and that there is likely dislocation of the humerus and glenoid which is related to a fractured and disassembled component of the humerus hardware.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPR 6MM HUM FRACT STEM PPS
Common Device Name
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9967236
Report Number0001825034-2020-01547
Device Sequence Number2282427
Product Code HSD
UDI-Device Identifier00880304591530
UDI-Public(01)00880304591530
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K140652
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number12-113556
Device Lot Number339720
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/26/2020
Supplement Date Received by Manufacturer08/07/2020
Initial Report FDA Received Date04/16/2020
Supplement Report FDA Received Date08/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
-
-