| Model Number |
21-7062-24 |
| Medical Device Problem Codes |
Material Integrity Problem (2978); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation results: one cadd extension set was returned for investigation in used condition.The sample was visually inspected at a distance of 12 to 24 inches and normal conditions of illumination. a cut was found in the tube.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.The investigator concluded that the cut in the tubing most likely occurred after the product left the manufacturing facility.The investigator noted that it was highly unlikely that the tubing was cut during the manufacturing process.The problem source of the reported product problem was unknown.A true root cause was not established.
|
| |
|
Event or Problem Description
|
|
It was reported that the extention line on the cadd extension set was severed.The patient went to hospice care with this product problem.No patient injury or complications were reported in relation to this event.
|
| |
|