Investigation summary: investigation flow: damage.Visual inspection: the helical blade inserter (p/n: 319.004, lot number: 5561095) was received at us cq.Upon visual inspection, the device is dented, nicked/scratched, and has deformations at the proximal end and inside the alignment indicator component.There was no evidence of breakage.Device failure/defect identified? yes.Dimensional inspection: no dimensional inspection was performed due to post-manufacturing damage.Document/specification review: no design issues or discrepancies were identified.Complaint confirmed? no.Investigation conclusion: this complaint is not confirmed as there is no evidence of breakage.However, the device has nicks, dents, and deformations.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part number: 357.372.Synthes lot number: 6442034.Supplier lot number: n/a.Release to warehouse date: jul 30, 2010.Expiration date: n/a.Manufactured by synthes brandywine.No ncrs were generated during production.Device history review: review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|