MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT2H08
Medical Device Problem Codes Material Frayed (1262); Failure to Advance (2524)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, a smart coil would not advance within its introducer sheath; therefore, it was removed.The procedure was completed using a new smart coil.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Please note that this complaint was submitted to the fda by the user facility with the following reference number: (b)(4).
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, a smart coil would not advance within its introducer sheath; therefore, it was removed.It was reported that the back end of the smart coil had frayed off.The procedure was completed using a new smart coil.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key9968462
Report Number3005168196-2020-00565
Device Sequence Number19757947
Product Code HCG
UDI-Device Identifier00814548015576
UDI-Public00814548015576
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/26/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2024
Device Model Number400SMTXSFT2H08
Device Catalogue Number400SMTXSFT2H08
Device Lot NumberF90172
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 03/26/2020
Supplement Date Received by Manufacturer01/14/2005
Initial Report FDA Received Date04/16/2020
Supplement Report FDA Received Date05/14/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-