ALLERGAN (COSTA RICA) INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number TRF-520
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Capsular Contracture (1761)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown, pain and concern with the product.
 
Event or Problem Description
Healthcare professional reported left side exchange due to patient's concern with the product, capsular contracture, baker grade unknown and pain.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified, crease fold, wear abrasion, the weight the device within specification and deformation.After autoclave cycle were observed: cloudy color in the gel and voids.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
 
Event or Problem Description
Healthcare professional reported left side exchange due to patient's concern with the product, capsular contracture, baker grade unknown and pain.Device has been explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9970135
Report Number9617229-2020-06411
Device Sequence Number11285880
Product Code FTR
UDI-Device Identifier10888628008366
UDI-Public10888628008366
Combination Product (Y/N)N
Initial Reporter StateUT
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/23/2021
Device Model NumberTRF-520
Device Catalogue NumberTRF-520
Device Lot Number2966205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2020
Initial Date Received by Manufacturer 03/20/2020
Supplement Date Received by Manufacturer05/26/2020
Initial Report FDA Received Date04/16/2020
Supplement Report FDA Received Date06/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-