MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR MMT-7020B ENLITE3 US; ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL Back to Search Results
Model Number MMT-7020B
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 04/12/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic the user reported inaccurate sensor readings that triggered a threshold suspend.The customer¿s blood glucose was 313 mg/dl and sensor glucose was 47 mg/dl at the time of the event, the difference is outside the acceptable range.Suspend on low limit in sensor settings was at 70 mg/dl.No harm requiring medical intervention was reported.The customer also reported getting calibration and change sensor errors.The sensor will not be returned for analysis.
 
Additional Manufacturer Narrative
Inspected one opened and used sensor and performed continuity resistance test and sensor passed per specifications.Performed bicarbonate buffer test sensor passed per specifications.This mdr related to the (b)(4) manufacturing site has been assigned a medwatch number from the medtronic minimed (b)(4) site, per variance 5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR MMT-7020B ENLITE3 US
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9970185
Report Number2032227-2020-104847
Device Sequence Number19395183
Product Code OZP
UDI-Device Identifier00763000140779
UDI-Public(01)00763000140779(17)210108(10)HG40MX3
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/08/2021
Device Model NumberMMT-7020B
Device Catalogue NumberMMT-7020B
Device Lot NumberHG40MX3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/23/2020
Initial Date Received by Manufacturer 04/13/2020
Supplement Date Received by Manufacturer04/29/2020
Initial Report FDA Received Date04/16/2020
Supplement Report FDA Received Date05/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age59 YR
Patient Weight200
-
-