• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 28MM -4 V40 TAPER VIT HEAD; HIP JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 28MM -4 V40 TAPER VIT HEAD; HIP JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS Back to Search Results
Model Number 6260-5-028
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/09/2014
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Lot specific voluntary recall was initiated for the v40 heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted, removed from the field and/or expired.
 
Event or Problem Description
Patient called stating she had a left hip replacement done (b)(6) 2014, she would like to know if her implants were involved in a recall.Patient is asymptomatic at the time of this call.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
28MM -4 V40 TAPER VIT HEAD
Common Device Name
HIP JOINT METAL/POLYMER SEMI-CONSTRAINED CEMENTED PROSTHESIS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9972630
Report Number0002249697-2020-00757
Device Sequence Number3775965
Product Code JDI
UDI-Device Identifier07613327012835
UDI-Public07613327012835
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2019
Device Model Number6260-5-028
Device Catalogue Number6260-5-028
Device Lot Number48781104
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/19/2020
Initial Report FDA Received Date04/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0261-2016
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age69 YR
Patient Weight72
-
-