• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. RELOAD- CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. RELOAD- CURVED CUTTER STAPLER; STAPLE, IMPLANTABLE Back to Search Results
Model Number CR40B
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch # p58y8h.Device analysis: the analysis results showed that the cs40b device was received with no apparent damage and with a cr40b reload loaded in the device.The reload was received with all staples, the washer uncut and with the knife recessed below the reload deck.The drivers were noted to be partially advance.The device was tested for functionality with a test reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.The condition of the reload is consistent with the device being partially activated.As a result of the partial firing the lockout was engaged when the device was reopened.Do not fire the instrument unless the closure trigger is properly latched against the handle.The firing trigger must be pulled back completely against the closure trigger to properly fire the instrument.In addition, if the firing sequence is not complete, you could deploy the staples without cutting the washer and forming the staples.Please reference instructions for use for additional information.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during an open gastric tube formation, a device loading the cr40b could not be fired at the 2nd firing on the stomach.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELOAD- CURVED CUTTER STAPLER
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
MDR Report Key9972637
Report Number3005075853-2020-02230
Device Sequence Number2784686
Product Code GDW
UDI-Device Identifier10705036001423
UDI-Public10705036001423
Combination Product (Y/N)N
PMA/510(K) Number
K040038
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2022
Device Model NumberCR40B
Device Catalogue NumberCR40B
Device Lot NumberR40074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/01/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/03/2020
Initial Report FDA Received Date04/17/2020
Patient Sequence Number1
-
-