| Model Number |
1601 |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
Therapeutic Response, Decreased (2271)
|
| Date of Event |
03/19/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Patient was having connection issues with the epg (trial stimulator) and percutaneous extension during the first few days of the trial period.In order to resolve the issue, an additional strain relief was added and the epg was replaced as there was concern it had become compromised.There has been no issues with the connection since the addition of the strain relief and replacement of the epg.
|
| |
|
Event or Problem Description
|
|
See section h, number 6, event problem and evaluation codes.
|
| |
|
Event or Problem Description
|
|
See section h, number 6, event problem and evaluation codes and section h, number 10, additional manufacturer narrative.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history records found no non-conformances.
|
| |
|